HMS-9700 Hall Effect Measurement System
Fully integrated Hall effect platform for measuring resistivity, carrier concentration, and Hall mobility from 4 K to 800 K, with fields up to 2 T (electromagnet) or 9 T (superconducting option).
Model: EM-7T
Water-cooled 7-inch electromagnet delivering up to 3.8 T, paired with the PS-6000 bipolar 4-quadrant power supply for precision field control in Hall effect, VSM, and magneto-optical setups.
Highlights
Interchangeable pole caps for gap/field trade-off
True 4-quadrant bipolar supply with smooth zero crossing
Closed-loop field control with NMR or Hall-probe feedback
Integrated water-flow and interlock protection
Vertical or horizontal field orientation with rotating base option
Technical Data
| Parameter | Value |
|---|---|
| Peak field | 3.8 T @ 10 mm gap; 2.0 T @ 40 mm gap |
| Pole diameter | 175 mm (7") |
| Field homogeneity | ±0.1% over 10 mm DSV (10 mm gap) |
| Field stability | ±5 ppm/h (closed-loop mode) |
| Power supply | PS-6000, ±60 A / ±80 V, 4-quadrant |
| Current accuracy | ±0.01% FS, 16-bit setting resolution |
| Sweep rate | Up to 1 T/min |
| Cooling | Deionized water, 8 L/min @ 4 bar |
| Weight | 380 kg (magnet), 45 kg (supply) |
Use Cases
Questions
Yes. The PS-6000 provides analog field programming input, RS-232/Ethernet interfaces, and TTL triggers, so it integrates with lock-in amplifiers, DMMs, and custom DAQ systems from any vendor.
Deionized water with resistivity above 1 MΩ·cm. A closed-loop chiller with DI cartridge is available as an accessory for labs without a DI water loop.
Full specifications and ordering information for the EM-7T.
Interested in this product? Tell us about your research needs — our engineers typically respond within one business day.
Explore
Fully integrated Hall effect platform for measuring resistivity, carrier concentration, and Hall mobility from 4 K to 800 K, with fields up to 2 T (electromagnet) or 9 T (superconducting option).
Research-grade VSM with 5×10⁻⁷ emu sensitivity, ±3 T electromagnet, and temperature options spanning 2 K to 1000 K for magnetic characterization of films, powders, and bulk materials.
Benchtop Hall-effect gaussmeter with 0.01 G resolution, DC to 30 kHz bandwidth, and temperature-compensated transverse/axial probes for field mapping and magnet QC.